Skip to content
English
  • There are no suggestions because the search field is empty.

CI-202 File Create Procedure

Video describing how to create a file on the CI-202

Purpose and Scope

This article describes how to create the first file on a CI-202 Create a File When the START/ENTER key is pressed, “Name: Files” is displayed on the top line, indicating the unit is in create files mode. The top line of the screen will read “Name” on the left, and a file name (e.g., “file00”) on the right. The bottom line will read “enter” at the left, “erase” at the center, and the four direction arrows at the right. Pressing the arrow keys will allow an alpha or numeric selection for a six-character file name. The right and left arrow keys select which character in the file name to edit and the up and down arrow keys are to choose a character. After naming the created file, press the START/ENTER key twice to continue the file creation process video showing how to create file The CI-202 must have a data file created in the file system for the unit to perform a measurement. If no files are present, the system will lock up when attempting to create a measurement. To clear the lock up, the user must remove the side cover then remove and replace the battery connection. This process is described in the video. Units are shipped from the factory with one file created on the file system. Please see attached video.

https://cidbio.sharepoint.com/:v:/s/Support/ETUpLDzoqK9GkjAZXpVKMtEBIs6hjhwzd2xgjaG5EM-dlw?e=6vYR2C