When exporting CI-110p data the flash drive needs to be less than half full

CI-110P stores all measurement data and images on the flash drive from USB port naturally, so that the user can view data from the instrument. When the user exports data from the CI-110P, the instrument creates a new Export_xxxx folder and Inside the Export_xxxx folder, all images and data will be duplicated and there is also a CSV file that can be viewed in Excel for the customer to analyze the data. Example: If the user's 128GB flash drive was more than half full of data/images when the export is started, there is not sufficient space on the 128GB flash drive to complete the process. Since the CI-110P effectively duplicates the data/images already on the 128GB flash drive in the instrument when the user exports data from the CI-110P, there are two recommendations: Since the user is gathering large data sets, they can use a larger capacity flash drive – perhaps 512GB The user should periodically backup the previous export folders to their computer so the export folder can be deleted from the flash drive. (If the user always leaves the exported folders on the flash drive it will eventually fill up again and the problem will reoccur)