CI-110P - no need to recalibrate after top display assembly replaced

The user does not need to recalibrate the instrument after the top display assembly (with Latte Panda board inside) is replaced in the instrument. The settings are saved in the instrument in the lower assembly, such that if the top display assembly is replaced. For example from a customer overseas, we can send a Disti a replacement Latte Panda board which when installed in the instrument, it will take the correct settings from the instrument. So this means the defective latte panda board inside a CI-110 can be fixed remotely and the whole instrument actually doesn't need to be returned for Latte Panda board replacement at our facility CI-110P top display assembly (with a LattePanda board inside).